RADWAG MYA 11/52.5Y ELLIPSIS SERIES DUAL RANGE MICROBALANCE, 11 G / 52 G X 1 / 10 MICROGRAM

RADWAG MYA 11/52.5Y ELLIPSIS SERIES DUAL RANGE MICROBALANCE, 11 G / 52 G X 1 / 10 MICROGRAM

 

OVERVIEW

Radwag MYA 11/52.5Y Ellipsis Series Dual Range Microbalance

The Radwag MYA 11/52.5Y Ellipsis series features some of the most unique and innovative features available in the balance market. It offers high quality automatic internal calibration, hands free operation the IR sensors, and an unmatched security of data. The 5Y is compliant in a variety of relevant functions and markets and offers connectivity to everything from a smart phone, tablet, or even a printer.

ELLIPSIS monitors temperature, humidity, pressure, and vibration. The results are displayed as graphs or a widget on the home screen. Unsuitable conditions for the balance are signaled by DWA. And all of this is recorded in a dedicated database for the end user to reference. This function alone can save the operator valuable time and provide insight as to why a certain measurement may skew data.


Please note that for international transactions, customers will be responsible for import taxes, clearance, and duty taxes.

Features

  • Dual Range: 11 g / 52 g by 1 / 10 microgram
  • Automatic internal calibration that senses changes in ambient conditions to maintain accuracy
  • Large 10" color touchscreen with customizable home menu
  • IR sensors for hand free operation
  • Smart automatic reflex leveling system
  • Works with RFID tags
  • Up-to-date information on ambient conditions
  • Real-time statistics from a series of measurements
  • Tool free disassembly of the weighing chamber
  • USP compliant
  • Pharmacopoeia compliant
  • Full data security
  • Electronic signatures, records, and audit trails
  • Quick review library that shows the last 20 measurements
  • Ability to add notes via text or voice to a series of measurements for reporting

Weighing Applications

  • Weighing
  • Checkweighing
  • Animal Weighing / Dynamic Weighing
  • Parts Counting
  • Percent Weighing
  • Dosing
  • Formulation
  • Density Determination
  • Statistics
  • Formulations
  • Pipette Calibration
  • Differential weighing
  • SQC
  • Peak Hold
  • Mass Control

ELLIPSIS will connect you directly to:

  • a smartphone or a tablet (hotspot)
  • a spreadsheet (USB-C, Free Link)
  • a printer (Ethernet, Wi-Fi®, USB-A)
  • a computer (Ethernet, Wi-Fi®, USB-A)
  • a barcode scanner (USB-A)
  • any large screen (HDMI)
  • your PC software (Radwag communication protocol)
  • a keyboard and mouse

Datasheet

Metrological parameters
Physical parameters
Communication interface
Electrical parameters
Environmental conditions
Maximum capacity [Max]

11/52

 g

Minimum load

0.1

 mg

Readability [d]

1/10

 µg

Verification scale interval [e]

1

 mg

Tare range

-52

 g

Standard repeatability [5% Max]

1.5

 µg

Standard repeatability [Max]

10

 µg

Standard minimum weight (USP)

3

 mg

Standard minimum weight (U=1%, k=2)

0.3

 mg

Permissible repeatability [5% Max]

2.7

 µg

Permissible repeatability [Max]

15

 µg

Linearity

±10/30

 µg

Eccentric load deviation

6/10

 µg

Stabilization time

max 10

 s

Adjustment

internal (automatic)

 

OIML Class

I

 

Leveling system

automatic - Reflex Level System

 

Display

10″ touchscreen

 

Delivery components

Microbalance, terminal, weighing pan, weighing pan shield, glass lid, power supply, pincette, brush, fabric dust cover.

 

Weighing chamber dimensions

ø90×90

 mm

Weighing pan dimensions

ø26 + ø40

 mm

Packaging dimensions

655×755×445

 mm

Net weight

9.1

 kg

Gross weight

16.6

 kg

Communication interface

USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot

 

Power supply

Adapter: 100 – 240V AC 50/60Hz 1A; 15V DC 2.4A
Balance: 12 – 15V DC 1.4A max*

 

Operating temperature

+10 — +40

 °C

Operating temperature change rate

±0.3°C/1h (±1°C/8h)

 

Relative humidity

40% — 80%

 

Relative humidity change rate

±1%/h (±4%/8h)