© Copyright 2022 by J&K Innovation. All Rights Reserved

RADWAG XA 41/120.5Y.A Analytical Balance 0.002 / 0.005 mg
OVERVIEW
Do you need a balance that not only determines weight with impressive accuracy, but also performs a digital audit, responds to your voice commands and sends you light messages?
If so, learn about the functionality of the ELLIPSIS series of balances:
Digital Weighing Auditor
Ensures that your balance is ready for use. It enables air buoyancy compensation in real-time. What is more, it informs about the need for a balance inspection or a periodic audit of the balance’s accuracy and sensitivity.
10-Inch Display
Bigger screen – 10’’ – means more information in one place. On the new ELLIPSIS terminal, you will see up to 3 screens, widgets, graphs, statuses, notes, handy measurement history.
Uncompromising User Verification
For the first time ever, the balance will verify the user using a password, RFID card, fingerprint reader, facial recognition mechanism, or any combination of these.
Ambient Light – An Innovative Way for the Balance to Communicate With the User
The light colour will tell you about status, process results, procedures, or alerts.
Hotspot
Use the balance on any device connected to it. This can be a smartphone, tablet, or computer.
RFID
ELLIPSIS works with RFID tags* that can be used to tag your products or formulation ingredients in the database. They can also be used to identify a user.
Live Note – Note Down Your Conclusions
It’s the first time you can add a voice memo to a measurement series or procedure report. If you prefer to write it down, ELLIPSIS allows you to add a text note.
Handy Library
This is where a series of recent measurements or a report ready for digital signature, according to 21 CFR Part 11, can be found.
Widgets
Prepared to display what you need most at any given moment.
Automatic Leveling of the Balance ReflexLEVEL System
Instead of manually leveling your scale - level it automatically. The ReflexLEVEL System saves you time and gives you maximum precision in your measurements.
Smart Min Weight
The balance displays whether the value of the weight you want to weigh is below or above the minimum weight threshold.
Two-point adjustment
Adjacency, which takes place at two measurement points instead of one, improves measurement precision over the full weighing range. And what does this mean for you? More accurate measurement.
Please note that for international transactions, customers will be responsible for import taxes, clearance, and duty taxes.
Datasheet
Physical parameters
Construction
Communication interface
Electrical parameters
Environmental conditions
Maximum capacity [Max] |
41 / 120 g |
Minimum load |
0.2 mg |
Readability [d] |
0.002 / 0.005 mg |
Verification unit [e] |
1 mg |
Tare range |
-120 g |
Standard repeatability [5% Max] |
0.004 mg |
Standard repeatability [Max] |
0.012 mg |
Standard minimum weight (USP) |
8 mg |
Standard minimum weight (U=1%, k=2) |
0.8 mg |
Permissible repeatability [5% Max] |
0.008 mg |
Permissible repeatability [Max] |
0.02 mg |
Standard linearity |
0.03 mg |
Permissible linearity |
0.1 mg |
Standard eccentric load deviation |
0.04 mg |
Permissible eccentric load deviation |
0.1 mg |
Sensitivity time drift |
1×10⁻⁶/Year×Rt
|
Stabilization time |
3.5 s |
Adjustment |
internal (automatic)
|
OIML Class |
-
|
Leveling system |
automatic – Reflex Level System
|
Display |
10″ graphic colour touchscreen
|
Weighing chamber |
automatic
|
Weighing chamber doors |
automatic
|
Delivery components |
Analytical Balance, weighing pan, weighing pan shield, centring ring, brush, fabric dust cover, power supply.
|
Weighing chamber dimensions |
200×170×220 mm |
Weighing pan dimensions |
ø64 mm |
Packaging dimensions W x D x H |
750×492×595 mm |
Net weight |
14.7 kg |
Gross weight |
16 kg |
Protection class |
IP 43
|
Communication interface |
2×USB-A, USB-C, RS 232 (COM3), HDMI, Ethernet, Wi-Fi®, Hotspot
|
Power supply |
Adapter: 100 – 240V AC 50/60Hz 1A; 15V DC 2.4A
|
Operating temperature |
+10 – +40 °C |
Operating temperature change rate |
±0.3 °C / 1 h (±1 °C / 8 h)
|
Relative humidity |
20% – 80%
|
Relative humidity change rate |
±1% / h (±4% / 8 h) |